Please use this identifier to cite or link to this item: http://www.repositorio.cdtn.br:8080/jspui/handle/123456789/416
Title: Absorbed doses on patients undergoing tomographic exams for pre-surgery planning of dental implants
Title of periodic: Applied Radiation and Isotopes New York
Authors: Zenóbio, Madelon Aparecida Fernandes
Silva, Teógenes Augusto da
Affiliation: Centro de Desenvolvimento da Tecnologia Nuclear/CDTN, Belo Horizonte, MG, Brasil
Centro de Desenvolvimento da Tecnologia Nuclear/CDTN, Belo Horizonte, MG, Brasil
Issue Date: 2007
Keywords: Patients;dentristry;tomography;radiation doses
Abstract: The thermoluminescent (TL) dosimetry was used to measure entrance skin absorbed doses at anatomical points close to critical organs of patients undergoing tomographic techniques as part of a pre-surgery planning for dental implants. The dosimetric procedure was applied in 19 patients, and absorbed doses could be measured with a combined uncertainty down to 14 percent. Results showed that patient doses may be increased by a factor of 20 in the helical computed tomography compared to panoramic and spiral conventional tomographic exams.