Use este identificador para citar ou linkar para este item: http://www.repositorio.cdtn.br:8080/jspui/handle/123456789/595
Título: Measurement of conversion coefficients between free in air kerma and personnel dose equivalent for diagnostic x-ray beams
Título do periódico: Radiation Measurement
Autor(es): Rosado, P.H.G
Nogueira, Maria do Socorro
Afiliação: Centro de Desenvolvimento da Tecnologia Nuclear/CDTN, Belo Horizonte, MG, Brasil
Centro de Desenvolvimento da Tecnologia Nuclear/CDTN, Belo Horizonte, MG, Brasil
Data do documento: 2008
Palavras-chave: Kerma;radiation doses;dose equivalents;thermoluminescent dosimetry
Resumo: Conversion coefficients between free in air kerma and personal dose equivalent were experimentally determined for the diagnostic X-ray qualities recommended by the International Electrotechnical Commission (IEC) for primary beams (RQR) and three RQA. Harshaw LiF-100H thermoluminescent dosemeters (TLD) were used for measurements after being calibrated against an ionization chamber traceable to the National Metrology Laboratory. A 300mmž300mmž150mm polymethilmethacrylate (PMMA) slab phantom was used for deep-dose measurements. The dosemeters were placed in the central axis of the X-ray beam at five different depths in the phantom (5, 10, 15, 25 and 35 mm) upstream the beam direction. The typical combined standard uncertainty of conversion coefficient value was 12
.
Acesso: L
Aparece nas coleções:Artigo de periódico

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
Art-01_MariadoSocorro_Nogueira.pdf303.04 kBAdobe PDFVisualizar/Abrir


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.