Título: | Determination of true null electrode spacing of an extrapolation chamber for x-ray dosimetry |
Título do periódico: | Journal of Physics: Conference Series |
Autor(es): | Figueiredo, M.T. T. Bastos, F. M. Silva, Teógenes Augusto da |
Afiliação: | Centro de Desenvolvimento da Tecnologia Nuclear, CDTN, Belo Horizonte, MG, Brasil |
Data do documento: | 2016 |
Palavras-chave: | Absorbed radiation doses;Calibration standards;Dosimetry;Electrodes |
Resumo: | An accurate determination of the true null distance between electrodes, that defines the sensitive volume of an extrapolation chamber, is critical for the establishment of a primary measurement method for absorbed dose in tissue. In this work, a critical analysis of two methodologies for determining the true null electrode spacing of a 23392 Böhm model PTW extrapolation chamber was done. |
Acesso: | L |
Aparece nas coleções: | Artigo de periódico |
Arquivos associados a este item:
Arquivo | Descrição | Tamanho | Formato | |
---|---|---|---|---|
Determination of true null electrode spacing of an extrapolation chamber.pdf | 1.2 MB | Adobe PDF | Visualizar/Abrir |